Overestimating DNA Amounts by UV Spectrophotometry? Consider the Plastic
Jun 25 2010
Microfuge tubes are one of the most commonly used pieces of labware in molecular biology labs around the globe. While great for holding and storing small samples, the chemicals used in the manufacturing of the
plastic could be leading to measurement errors for researchers. Kevin Lewis and his colleagues from Texas State University in San Marcos, TX describe their findings [1] that indicate that UV light–absorbing chemicals
are capable of leaching from polypropylene microtubes into samples, leading to erroneous measurements of nucleic acid concentration.
The authors surveyed a series of commercially available polypropylene microfuge tube brands under various experimental conditions and found that leaching occurred among all tubes examined.
Using mass spectrometry they identified a complex mixture of small molecular weight chemicals that absorbed UV light strongly at 220 and 260nm, wavelengths normally used to detect and measure protein and DNA concentrations. The leached chemicals resulted in concentration overestimates up to 300%.
“It was quite surprising that leaching of chemical additives which absorb light at the same wavelengths as DNA and proteins had not been described before because it is so ubiquitous among commercially available plastic tubes and biochemists have been using these smallplastic tubes for many decades,” said Lewis.
Although Lewis and his co-author Gary Beall do not have a solution for the problem, they are encouraged by recent trends in microtube manufacturing. “Some manufacturers have recently begun offering tubes that contain a reduced number of additives,” noted Lewis. For studies involving analysis of biological samples by absorbance spectroscopy, Beall and Lewis conclude that at the moment the use of polypropylene tubes with less additives appears to be the best way to reduce measurement errors due to leaching.
[1]. Interference with spectrophotometric analysis of nucleic acids and proteins by leaching of chemicals from plastic tubes L. Kevin Lewis, Michael H. Robson, Yelena Vecherkina, Chang Ji, and Gary W. Beall BioTechniques, Vol. 48, No. 4, April 2010, pp. 297–302
Digital Edition
LMUK 49.7 Nov 2024
November 2024
Articles - They’re burning the labs... Spotlight Features - Incubators, Freezers & Cooling Equipment - Pumps, Valves & Liquid Handling - Clinical, Medical & Diagnostic Products News...
View all digital editions
Events
Nov 11 2024 Dusseldorf, Germany
Nov 12 2024 Cologne, Germany
Nov 12 2024 Tel Aviv, Israel
Nov 18 2024 Shanghai, China
Nov 20 2024 Karachi, Pakistan