Mass Spectrometry & Spectroscopy
Highly Analytically-Sensitive Analysis of Aldosterone in Plasma on the Xevo TQ-XS for Clinical Research
Jul 04 2016
Author: Dominic Foley on behalf of Waters Corporation
Goal
To demonstrate the analytical sensitivity gains obtained for the analysis of aldosterone on the Xevo® TQ-XS Mass Spectrometer compared to the Xevo TQ-S Mass Spectrometer for clinical research
Background
Aldosterone is a mineralocorticoid steroid hormone produced by the zona glomerulosa of the adrenal cortex. The production of aldosterone is assessed during pharmacological clinical research studies of aldosterone synthase (CYP11B2) inhibitors. It can be present at low concentrations in plasma (<25 pg/mL), making it difficult to measure with an acceptable level of precision and accuracy.
The Xevo TQ-XS (Figure 1) is an advanced benchtop, tandem quadrupole mass spectrometer expanding the scope of ultimate analytical sensitivity analysis. It features StepWave™ XS ion transfer optics, delivering enhanced analytical sensitivity, robustness, and reliability. In addition, the new Xtended Dynamic Range (XDR) Detector allows six orders of linear dynamic range. This ensures that methods are easily transferable onto the instrument and wide concentration ranges can be measured in a single analysis.
Figure 2. Injection of a 1 pg/mL solution (n=3, 50 fg on-column) of aldosterone on the Xevo TQ-XS.
Solution
In this technology brief, the Xevo TQ-XS was used to significantly improve the analytical sensitivity and imprecision of a LC-MS/MS method for the analysis of aldosterone in plasma.
Sample Preparation and LC-MS/MS Analysis
Plasma samples were precipitated with zinc sulphate/methanol and diluted with phosphoric acid before loading onto the Oasis® MAX μElution Plate. Consecutive washes with phosphoric acid, ammonia in 10% methanol, and water were performed. Samples were eluted with 70% aqueous methanol and diluted with water.
Samples were injected on an ACQUITY UPLC® I-Class System utilising a water/methanol gradient and a CORTECS C18 column. The same UPLC system and solvents were used to evaluate both the Xevo TQ-S and Xevo TQ-XS.
MRM transitions monitored for aldosterone were 359.2 > 189.0 (quantifier) and 297.2 (qualifier). The internal standard – aldosterone-4H2 – had a MRM transition of 363.2 > 190.0.
To convert conventional mass units (pg/mL) to SI units (pmol/L) multiply by 2.774.
Results
A system comparison using a 1 pg/mL solution of aldosterone was performed to demonstrate the analytical sensitivity improvement observed for aldosterone on the Xevo TQ-XS (Figure 2). This analytical sensitivity improvement also contributes to a significant improvement in imprecision of the peak area at this concentration (<2% RSD, n=3).
The low calibration serum standard used for the LC-MS/MS method was assessed on each system over three separate days. This evaluation demonstrates reproducibility of a sample extracted and injected on both the Xevo TQ-XS and Xevo TQ-S, on three separate occasions (Figure 3).
An evaluation using an extracted aldosterone plasma sample was performed, demonstrating the analytical sensitivity improvement on the Xevo TQ-XS that was observed for analysing endogenous aldosterone in plasma (Figure 4).
Total imprecision of the LC-MS/MS method was assessed on both the Xevo TQ-XS and Xevo TQ-S over three separate days, with ten replicates on each day at 21, 42, 174, and 669 pg/mL (n=30). Total imprecision was shown to improve on the Xevo TQ-XS (Table 1).
Table 1. Total imprecision comparison on the Xevo TQ-XS and Xevo TQ-S ranging from 21–669 pg/mL.
Summary
The Xevo TQ-XS has demonstrated significantly improved performance of our LC-MS/MS method for the analysis of aldosterone for clinical research.
The benefits of this system for the analysis of aldosterone include:
• Increased MS analytical sensitivity, signal-to-noise, and imprecision for aldosterone in solution down to 1 pg/mL (<2% RSD)
• Increased method analytical sensitivity for aldosterone in stripped serum at 15 pg/mL and aldosterone in plasma at 11.6 pg/mL (32 pmol/L)
• Improved imprecision (<8% CV) of the method at the four aldosterone QC concentrations in plasma.
Digital Edition
Lab Asia 31.6 Dec 2024
December 2024
Chromatography Articles - Sustainable chromatography: Embracing software for greener methods Mass Spectrometry & Spectroscopy Articles - Solving industry challenges for phosphorus containi...
View all digital editions
Events
Jan 22 2025 Tokyo, Japan
Jan 22 2025 Birmingham, UK
Jan 25 2025 San Diego, CA, USA
Jan 27 2025 Dubai, UAE
Jan 29 2025 Tokyo, Japan