Waters within power plants, and particularly boiler feed waters are subject to strict analytical control. Organic impurities in the boiler feed water can easily be oxidised under high pressure and high temperatures, and may lead to the formation of corrosive organic acids and carbonic acid. Salts are often added as oxygen scavengers.
TOC monitoring in the boiler feed water is crucial for controlling this process. This is a matter of trace contamination in the range below 1 ppm TOC.
Wet chemical UV digestion is particularly suited for the TOC determination within this concentration range and this matrix, not least due to its large volume of sample injection (up to 20 ml) and its high measurement sensitivity. ­

Experimental
A similar matrix as generally found in power plant waters was synthesised and defined TOC contents were added.
The following concentrations and substances were used for the synthetic matrix:
• 7 g/l boric acid
• 1 mg/l sodium
• 7 mg/l potassium
• 2 mg/l lithium
• 7 mg/l ammonium
A 300 µg/l and a 1000 µg/l TOC-standard (potassium hydrogen phthalate) were then added to this synthetic solution.
The synthetic samples were adjusted to pH 2 with 2N sulphuric acid.
The following method parameters were set at the instrument system.
Measurement parameters    NPOC
Digestion technology    Wet chemical UV digestion (185nm and 254nm)
Number of repeated measurements    3 of 4
Purging time    300 seconds
Injection volume    5000 µl
Additional reagents (oxidising agent)    No

A TIC control measurement ensured that the inorganic carbon and dissolved carbon dioxide had been eliminated.
An autosampler with parallel purging and analysis function was used for feeding the sample. This reduces analysis times and allows a high sample throughput.
The non-purgeable organic carbon (NPOC) proportion was oxidised to CO2 in the UV reactor without additional use of an oxidising agent and determined by means of an NDIR detector.
Calibration
The software of the analyser allows the combination of one measuring method with several calibration ranges. The most suitable calibration curve for calculating the results is automatically selected by the analyser based on the surface integrals. This ensures that the optimal calibration range is always selected. This is particularly user-friendly and guarantees highest accuracy and precision of the measured values. A single-point calibration with 1000 ppb TOC was prepared from potassium hydrogen phthalate for the synthetic samples for the field of power plants.

Results
The following table shows the mean values of the NPOC measurements as well as the relative standard deviation.


Summary
The results show that there is no need to add an oxidising agent charged with a TOC blank value such as sodium peroxodisulphate for the complete oxidation of waters in the field of power plants. The UV radiation and the dissolved oxygen in the sample are sufficient to convert all organic compounds into CO2. The results are extremely reliable TOC measurements in the boiler feed water, other ultrapure waters and even in drinking water analyses.

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

International Labmate 49.6 - Sept 2024

September 2024

Chromatography Articles - HPLC gradient validation using non-invasive flowmeters Mass Spectrometry & Spectroscopy Articles - From R&D to QC, making NMR accessible for everyone: Putting NMR...

View all digital editions

Events

ISC 2024

Oct 06 2024 Liverpool, UK

SCANAUTOMATIC

Oct 08 2024 Gothenburg, Sweden

AQE 2024

Oct 09 2024 Birmingham, UK

WWEM

Oct 09 2024 NEC, Birmingham, UK

SPICA 2024

Oct 15 2024 Milan, Italy

View all events