The production of bleach requires Chlorine, which is manufactured by the process of electrolysis. It is important to analyse the quality of the Chlorine gas to ensure that the electrolysis process is efficient with respect to time, therefore not wasting electricity and to ensure there are no leaks between the membranes. By analysing the impurities Hydrogen, Oxygen, Nitrogen and Carbon Dioxide in Chlorine gas will indicate the efficiency of the process and any leakage.

Ellutia have developed a customised version of the 200 Series GC with Multi Dimensional Chromatographic channels each with a TCD to perform this analysis.

Channel A is for the analysis of Hydrogen in the range 0.01 to 0.5% vol. using Argon carrier gas.

Channel B is for the analysis of Oxygen, Nitrogen and Carbon Dioxide in the range 0.01 to 3% vol. using Helium carrier gas.

Special Hastelloy valves, pipework and columns have been installed to reduce corrosion by the sample. Multi Dimensional Chromatography techniques are employed to ensure the Chlorine matrix gas does not reach the TCD. With the added heated Naffion dryer installed in the sample line (to remove moisture thus reducing potential corrosion) the analyser can be placed ‘on-line’.

For more information visit http://j.mp/UhhiXz


Digital Edition

International Labmate Buyers' Guide 2024/25

June 2024

Buyers' Guide featuring: Product Listings & Manufacturers Directory Chromatography Articles - Enhancing HPLC Field Service with fast-response, non-invasive flowmeters - Digital transformatio...

View all digital editions

Events

Asia Labex

Jul 03 2024 Gandhinagar, India

EuCheMS Chemistry Congress

Jul 07 2024 Dublin, Ireland

HPLC 2024

Jul 20 2024 Denver, CO, USA

ICMGP 2024

Jul 21 2024 Cape Town, South Africa

ADLM 2024

Jul 28 2024 San Diego, CA USA

View all events