Mass Spectrometry & Spectroscopy
Determination of Fat, Oil and Grease (FOG) from wastewater
Mar 17 2020
Author: Kevin Nott on behalf of Oxford Instruments NanoAnalysis
Free to read
Articles are free to download. Unlock the article to be shown more content, graphs and images.
A method to increase the speed and accuracy of Fat, Oil and Grease (FOG) analysis in wastewater has been developed. This simple, reliable technique is suitable for non-specialised personnel. It requires minimal sample preparation and results in increased throughput with automated analysis. Additionally, the technique is non-destructive, allowing solvent extraction verification after analysis.
Application
Fat, Oil and Grease (FOG) is often described as a pollutant since it is discharged in wastewater causing build-up and subsequent blockage of sewers. Therefore, it is a crucial parameter in environmental monitoring performed by water suppliers’ laboratories and environmental authorities. In the United States, ‘oil and grease’ is classed as a conventional pollutant under the US Clean Water Act. The US Environmental Protection Agency method 1664 (solvent extraction followed by gravimetry) is used for survey and monitoring programs related to this and other legislation. In the United Kingdom, the most commonly used method separates FOG from the water by passing it through a glass fibre filter (1.2μm pore size). The FOG is extracted from the filter using petroleum ether (leaving the solid particulates behind) with the total amount determined by gravimetry after the solvent has been boiled off [1]. In contrast to the labour-intensive and destructive solvent extraction method, this article describes a fast, accurate and solvent-free variant of the method whereby the FOG is directly measured on the filter without extraction.
Advantages of NMR
Solvent extraction is commonly used for the measurement of FOG deposited on a filter. However, this method can be time consuming, requires skilled operators and the use of hazardous solvents.
In contrast, NMR does not require hazardous solvents which are costly to dispose of. It is a bulk measurement technique which selectively measures all of the oil and fat in the sample regardless of the distribution of the grease on the filter. A linear calibration is readily generated using a few gravimetrically prepared samples. NMR enables rapid measurement of oils and fats with different fatty acid compositions.
The MQC+ benchtop NMR analyser provides an alternative to extensive wet-chemistry methods. FOG analysis with the MQC+ is quick and easy to perform, simple to calibrate and requires minimal sample preparation. As such it is ideal for non-specialist laboratory personnel and the whole process can be automated using the MQ-Auto autosampler.
Method
Benchtop NMR detects the signal from oil and fat on a single filter sheet after the initial NMR signals from the solids and any residual moisture have decayed completely. The remaining signal intensity correlates with the mass of FOG on the sample.
Calibration
Since NMR calibrations are always linear, only two well characterised standards are required to calibrate the analyser. However, it is recommended that the instrument is calibrated using at least 6 standard samples with known FOG contents, evenly distributed over the range of interest. The samples must be dry prior to the NMR measurement.
In this study, the calibration standards were prepared by adding 100 mg of the desired composition of fat, oil or grease (edible oil, animal fat, mineral oil, engine lubricant, etc.) to petroleum ether (100ml), followed by adding aliquots of this mixture to clean filter sheets. After drying, the reference grease content of each standard was measured using a precision analytical balance to at least four decimal places (± 0.1 mg).
Measurements
To prepare a sample for analysis, a dried filter sheet containing filtration residuals was folded and placed into a small glass vial. The samples were then temperature conditioned at 40°C in a dry block for a minimum of 20 minutes prior to measurement by the MQC+. The total measurement time for one sample was 5 minutes, excluding sample conditioning time.
A comparison of FOG content measurements performed on the calibration standards against gravimetry is shown in Table 1; it clearly shows that the NMR method is more accurate than solvent extraction (using petroleum ether).
Table 1. Comparison of solvent extraction and NMR results for FOG content against gravimetry
Results
Comparison of NMR and solvent extraction using artificial standards.
Figure 1 shows a calibration obtained by measuring a set of calibration standards created by adding known amounts of a FOG mixture to various filter papers.
Comparison of NMR and solvent extraction using wastewater samples
In addition, a set of real samples prepared by passing wastewater through filters were measured non-destructively by NMR (using the calibration produced from artificial standards in Figure 1) followed by the extraction method; the results are shown in Figure 2. The two methods show a very good correlation, except for sample 17 from which glycerol was also extracted (derived from soap-based products) and sample 20 which was an inhomogeneous sample. The data also shows that the limit of detection of the NMR method is 1mg/L (compared to 8mg/L for the solvent extraction method).
Comparison of NMR and solvent extraction repeatability using wastewater samples
A set of twelve filter papers were prepared using the same sample of wastewater. Again, these were analysed by NMR first followed by solvent extraction; the results are shown in Table 2.
The t and f tests applied to this data (95% confidence interval) show there is a significant difference between the means of the two methods but not the standard deviations; the NMR results are slightly higher than those for solvent extraction which tends to have a low recovery, Furthermore, it is noticeable that the results by the two methods correlate with each other for this composite sample.
Conclusions
The sensitivity of the MQC+ analyser enables accurate measurement of FOG on a single filter sheet; the limit of detection of the filtration followed by NMR method is 1mg/L. NMR measurement precision is comparable to the solvent extraction method, but potentially more consistent as it is less prone to operator error. Minimal sample preparation is required, removing the need to use hazardous solvents and other media. The single sample measurement is rapid (typically 5 minutes) compared to solvent extraction which can take a few hours. Samples can be measured automatically in batches with the MQ-Auto during the day and night. The NMR technique is non-destructive enabling convenient repeat measurements.
Further information: visit nmr.oxinst.com/mqc or email magres@oxinst.com
References
1. ‘The Determination of Oil and Greases in Waste Water by Filtration, Solvent Extraction and Gravimetry’ in Methods for the examination of waters and associated materials, 1987
(ISBN 011 752076 4)
Free to read
Articles are free to download. Please login to read this article or create an account.
Digital Edition
Lab Asia 31.6 Dec 2024
December 2024
Chromatography Articles - Sustainable chromatography: Embracing software for greener methods Mass Spectrometry & Spectroscopy Articles - Solving industry challenges for phosphorus containi...
View all digital editions
Events
Jan 22 2025 Tokyo, Japan
Jan 22 2025 Birmingham, UK
Jan 25 2025 San Diego, CA, USA
Jan 27 2025 Dubai, UAE
Jan 29 2025 Tokyo, Japan